Instructions for use - Tiva2 - Tiva2H Thermodisinfector
ALARM DESCRIPTION ACTION
E01: BLACKOUT Power failure during the cycle Press RESET
and the program blocked.
E02: DOOR OPEN The door is open or unlocked. Close it. Press RESET
If it persists, please contact a technician.
E02: DOOR OPEN Close it. Press RESET
If it persists, please contact
a technician.
E07: NO DOOR LOCK The door did not lock within the time Press RESET
preset. If it persists, please contact
a technician.
E09: NO DOOR UNLOCK The door did not unlock within the time Press RESET
preset. If it persists, please contact
a technician
E11: DIRTY WATER Check for dirt residues on the Press RESET
loadandcheckthepuriedwater Ifitpersists,pleasecontact
in the network. a technician.
E20: NO COLD WATER The cold water valve could be Press RESET
closed or partially open. Check it. If it persists, please contact
a technician.
E21: NO HOT WATER The hot water valve could be Press RESET
closed or partially open. Check it. If it persists, please contact
a technician.
E22: NODEMIWATER Thepuriedwatervalvecould PressRESET
be closed or partially open. If it persists, please contact
Check it. a technician.
E26: PRINTER PAPER The printer ran out of paper. Insert a Press RESET
new roll. If it persists, please contact
a technician.
E30: FLOWMETER 1 Tank 1 of the chemical liquid could Press RESET
be empty. Check it. If it persists, please contact
a technician.
E31: FLOWMETER 2 Tank 2 of the chemical liquid could Press RESET
be empty. Check it. If it persists, please contact
a technician.
E32: FLOWMETER 3 Tank 3 of the chemical liquid could Press RESET
be empty. Check it. If it persists, please contact
a technician.