Instructions for use - Tiva2 - Tiva2H Thermodisinfector
E66: NO HEAT There is a problem with the heating Press RESET
of the water. If it persists, please contact
a technician.
E67: CONDENSER L. The water level in the steam condenser Press RESET
has reached the maximum level allowed. If it persists, please contact
a technician.
E70: PUMP PRESSURE There may be water leaks. Press RESET
If it persists, please contact
a technician.
E71: HEPAFILTER TheHEPAltercouldbedirtyorobstructed. PressRESET
Check it. If it persists, please contact
a technician.
E75: PRODUCT 1 EMPTY Tank 1 of the chemical liquid is empty. Press RESET
Add liquid. If it persists, please contact
a technician.
E76: PRODUCT 2 EMPTY Tank 2 of the chemical liquid is empty. Press RESET
Add liquid. If it persists, please contact
a technician.
E77: PRODUCT 3 EMPTY Tank 3 of the chemical liquid is empty. Press RESET
Add liquid. If it persists, please contact
a technician.
E89: NO DISINFECTION The minimum value of A0 has not been reached Press RESET
and the load is still contaminated. If it persists, please contact
a technician.
Warning Description Action
1 RELOAD SALT BOWL 1-Remove the basket Press RESET
2-Loosen the salt bowl cap for 5 seconds
3-Fill the salt bowl
4-Tighten the salt bowl cap
2 MAINTENANCE REQUEST Contact customer service or the technician Press RESET
in charge of periodic maintenance.
3 VALIDATION REQUEST Contact customer service or the technician Press RESET
in charge of periodic validation.