Troubleshooting
108
12.5.2 Error/status messages
Int. pressure
Alarm in the event of inadequate combustion gas or calibration gas.
Adjustable threshold; default: 14 mbar
Overtemp
The flame is too hot. Wobbe signal > 76 mV
Temperature rise in the inner tube of the thermoelectric battery > 50°C
Fault
Gas pressure is too low.
Fault in the thermoelectric battery.
Defective PT 100 temperature sensor.
Service
Air inlet temperature > 52°C
Operation
Flame is burning, and calorimeter is running on combustion or calibration gas.
Operation delay
Flame is burning, calorimeter is running on combustion or calibration gas.
Delay time has elapsed, after which the analog signals are enabled.
Process
Flame is burning, and solenoid valve for calibration gas is open.
Calorimeter is running on combustion gas.