Safety instructions
3.7.1 Information on explosion protection
When opening, observe the instructions on explosion protection in hazardous
areas! May be opened by specialists only!
The device must first be flushed after opening the housing and switching on
again! Minimum flow volume is 16,000 l/h of compressed air at 4.5 – 5 bar!
You must wait for at least 5 minutes after separation of the power supply before
any doors of the housing may be opened.
Static electricity discharge! Warning plate may only be cleaned with a damp cloth!
3.8 Owner safety precautions
• Toxic gases – The owner must provide suitable protective equipment for the
combustion calorimeter that can reliably prevent injuries to personnel, e.g.
from escaping gas!
• Tripping hazard from improperly routed supply lines!
Other owner-side safety precautions:
Corresponding section!