Transport, installation and acceptance
51
Transport securing device in the pressure controller
Fig. 6.9: Transport securing devices of pressure controller
Transport securing device
Order of removal/attachment of
transport securing device
Transport securing device of pressure
controller
Remove the retaining belts.
Open the protective flap (screw
closure).
Remove the transport securing device
(bubble wrap).
Reclose the protective flap.
Proceed in reverse order to reattach the transport securing device.
Transport securing device of density measuring cell
Transport securing device
Order of removal/attachment of
transport securing device
Transport securing device of density
measuring cell
Remove the white cable ties around
the density measuring cell.
Loosen the screw.
Density measuring cell must swing
freely.
Proceed in reverse order to reattach the transport securing device.
Observe the length of the screw!