Transport, installation and acceptance
36
6.3.2 Wall mounting
WARNING
Risk of injury due to the weight of the device!
For weight, see technical data!
Specify measures to prevent the device from falling down and use suitable
hoisting devices!
The combustion calorimeter is intended for mounting to a solid wall. There are four
11 mm fastening eyelets on the housing . The wall used for mounting must be
sufficiently sturdy to bear the weight of the combustion calorimeter.
Adequate distance must be provided between the side wall of the installation
location and the combustion calorimeter for service and maintenance (left side:
approx. 60 cm, right side: approx. 50 cm).
Fig. 6.1: Wall mounting