Main menu - Options - Calibration - Automatic calibration
The automatic calibration is configured in this menu.
The following must be observed here:
Input:
Day is a list field (Su, Mo, Tu, We etc.). Time and Cycle are value fields.
Duration of calibration:
Depending on the device type, the calibration duration is 10 20 min.
NOTE
With a restart of the device or a change of calibration gas, the calibration may
have to be started multiple times until the pulse line is sufficiently flushed with
calibration gas and the stability criterion is met.