Transport, installation and acceptance
43
6.3.5 Electrical connection
DANGER
Risk of electric shock!
Changes to the electrical equipment of the combustion calorimeter may be
carried out only by skilled electricians in accordance with the electrotechnical
rules.
Parts of the open combustion calorimeter marked with the adjacent symbol may
still carry voltage even when the main switch is switched off! If required,
separate the combustion calorimeter from the supply network!
DANGER
Risk of explosion!
Non-approved cable bushings pose a hazard! Cable bushings which are not
suitable for the cable cross section pose a hazard!
Use only cable bushings with ATEX/IECEx approval which fit the cable cross
section!
Explosion protection is no longer guaranteed if seals are damaged or doors are
closed improperly!