Transport, installation and acceptance
37
6.3.3 Gas connections
DANGER
Risk of explosion!
Non-approved lines and insufficient connections pose a hazard!
Gas connections may only be installed by qualified personnel!
Check for proper sealing of connections and ensure this during operation!
NOTE
Accessories must be clean and free of residue. Contaminants may reach
inside the combustion calorimeter and may cause false measurements
and/or damage.
The input pressure for the gas connections must not exceed the
specifications of the technical data of the combustion calorimeter.
Check each connection carefully for leaks. When leaks are present, the
system draws air and indicates false measured values.
Do not use sealant for sealing the gas connections. Sealant components
may falsify the measurement result.
Only suitable lines may be used.
6.3.3.1 Ignition protection gas supply Compressed air
Compressed air is required as the ignition protection gas for operation of the
combustion calorimeter. Ensure sufficient supply and seal of the connections
check connection point for proper seal.
Device requires compressed air at 4.5 5 bar, 16,000 30,000 l/h, for operation.
Air recommendation as per ISO 8573-1, class 533 (solid material: 40 μm (class
5)/dew point: 20°C (class 3)/oil quality: 1 mg/m³ (class 3))
NOTE
When installing a shut-off valve of the ignition protection gas supply
Post warning information: Warning Valve of the ignition protection gas
supply Follow instructions before closing!