Transport, installation and acceptance
39
6.3.3.3 Carrier gas supply
NOTE
For combustion gases that do not have stable combustion, a carrier gas can
be added.
Carrier gases maintain combustion and make no increased contribution to
the combustion value of the combustion gas.
Combustion calorimeters can be subsequently converted to a carrier gas
supply. Contact the manufacturer regarding this.
6.3.3.4 Calibration gas
WARNING
If a pressure reducer is not installed, escaping calibration gas must be
discharged by the owner to a safe environment!
NOTE
Calibration gas must be free of contaminants and condensate.
The calibration gas quality must be similar to the combustion gas quality
(see Section 1.10).
For device versions with two measuring ranges, up to two calibration gases
may be necessary.
Software-side configuration of the calibration gas.
(
Section 9.3 Available displays)