5-5. Zero Tracking (ZT)
・Zero tracking clear Final to zero at every specified period, if zero shift
during that period remains within the specified range.
・Allowed tracking period is from 0.0 to 9.9 seconds, and Zero
tracking range is from 0 to 9999 in 1/4 of scale division unit.
(0002=0.5 divisions, 0012=3 divisions)
Zero tracking is disabled if zero is specified for the period or
range.
Boundary of zero track
Tracking period
Weighing value
Tracking range
+
0
-
Tracking period
From the point when it returned within the range,
counting will be resumed.