Compensation setpoints
821 0334
hex
Temperature
compensation setpoint. If
a value is written to this
register, probe uses it
instead of its own
temperature
measurement.
When probe heating is in
use, temperature must be
written to this register to
enable output parameters
that depend on
temperature measurement
(for example, relative
humidity).
32-bit float Unit: °C
0335
hex
Purge
773 0304
hex
Purge interval 32-bit float Unit: min
10 ... 14400
0305
hex
1283 0502
hex
Interval purge on/o 16-bit
boolean
0 = O
1 = On
1284 0503
hex
Startup purge on/o 16-bit
boolean
0 = O
1 = On
Filtering
795 031A
hex
Measurement filtering
factor
32-bit float Range: 0.000 ... 1.000
1.000 = Reading shows
100% of the most recent
measured value (no
filtering, default)
0.01 ... 0.99 = Reading
shows 1 ... 99% of the most
recent measured value and
part of the previous
reading (filtering is
applied). For example,
"0.9" means that the
filtered measurement
reading = 90% of the most
recent measured value +
10% of the previous
reading.
031B
hex
Appendix A – Modbus reference
83