Filtering
1282 0501
hex
Enable of disable
measurement filtering
using the user-defined
filtering factor (register
031A
hex
)
16-bit
boolean
0 = O
1 = On
Communication
1537 0600
hex
Modbus address 16-bit
integer
1 ... 247
Default: 240
1538 0601
hex
Bit rate enum 0 = 300
1 = 600
2 = 1200
3 = 2400
4 = 4800
5 = 9600
6 = 19200
7 = 38400
8 = 57600
9 = 115200
1539 0602
hex
Parity, data, stop bits enum 0 = None, 8, 1
1 = None, 8, 2
2 = Even, 8, 1
3 = Even, 8, 2
4 = Odd, 8, 1
5 = Odd, 8, 2
(default: 1 = None, 8, 2)
1540 0603
hex
Response delay 16-bit
integer
Unit: ms
Range: 0 ... 1000
Functions
1285 0504
hex
Start purge 16-bit
function
status
When writing to register:
1 = Start purge
1542 0605
hex
Restart device 16-bit
function
status
When writing to register:
1 = Restart the device
HMP3, HMP4, HMP5, HMP7, HMP8, HMP9, MMP8, TMP1 User Guide M212022EN-D
84