EasyManua.ls Logo

Vaisala HUMICAP HMT310 - FILT Set Result Filtering; RESET Reset Transmitter; ERRS Display Error Messages

Vaisala HUMICAP HMT310
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
USER'S GUIDE____________________________________________________________________
50 __________________________________________________________________ M210619EN-D
FILT Set Result Filtering
Syntax: FILT xx<cr>
Enable or disable the filtering or select the extended filter to reduce noise
of the measurement.
Where
x = ON, OFF or EXT
ON = Short filter of about 15 s (results the average value of the
last 15 s measurement data)
OFF = No filtering (default)
EXT = Extended filter of about 1 min (results the average value
of the last 1 min measurement data)
RESET Reset Transmitter
Syntax: RESET<cr>
ERRS Display Error Messages
Syntax: ERRS<cr>
Display transmitter error messages. If there are no errors present,
a PASS will be returned.
Examples:
>errs<cr>
PASS
>
>errs<cr>
FAIL
Error: Temperature measurement malfunction
Error: Humidity sensor open circuit
>

Table of Contents

Related product manuals