Chapter 4 ________________________________________________________________ Operation
VAISALA_______________________________________________________________________ 39
Chemical purge
PUR [ON/OFF] Enable or disable automatic
chemical purge
PURGE Activate chemical purge
PURR [ON/OFF] Enable or disable automatic
chemical purge in power up
Sensor heating
XHEAT [ON/OFF] Enable or disable sensor heating
Measurement Output
R Start Continuous Outputting
Syntax: R<cr>
Starts output of measurements to the peripheral devices (RUN-mode); the
only command that can be used is S (stop).
The output mode can be changed with command FORM.
Example:
>r<cr>
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C
S Stop Continuous Outputting
Syntax: S<cr>
Stops the continuous output. Also the Esc key (on your computer
keyboard) can be used to stop outputting.