Chapter 4 ________________________________________________________________ Operation
VAISALA_______________________________________________________________________ 51
Setting, Scaling and Testing Analog Outputs
AMODE Set Analog Outputs
(0/4 ... 20 mA)
Syntax: AMODE ch1 ch2<cr>
Where
ch1 and ch2 = I0 = 0 ... 20 mA
I1 = 4 ... 20 mA
Example:
>amode i1 i1<cr>
Ch1 output mode: 4...20mA
Ch2 output mode: 4...20mA
>
ASEL Select Parameter for Analog
Outputs
Syntax: ASEL xxx yyy<cr>
Where
xxx = Quantity of channel 1
yyy = Quantity of channel 2
Use abbreviations shown in Table 9 on page 52.