USER'S GUIDE____________________________________________________________________
64 __________________________________________________________________ M210619EN-D
1. Remove the filter from the probe.
2. Remove the damaged sensor and insert a new one. Handle the new
sensor by the plastic socket.
CAUTION
Do not touch the sensor plate.
3. After sensor change the humidity calibration must be made
according to the instructions, see FCRH Relative Humidity
Calibration after Sensor Change on page 68.
4. Install a new filter on the probe. When using the stainless steel
filter, take care to tighten the filter properly (recommended force is
130 Ncm).
New sensors can be ordered from Vaisala, see Parts List for
Consumables below.
Parts List for Consumables
Table 10 Available Spare Parts
Spare Part Order Code
PPS Plastic Grid with Stainless Steel Netting DRW010281SP
PPS Plastic Grid Filter DRW010276SP
Sintered Filter AISI 316L HM47280SP
Stainless Steel Filter HM47453SP
Stainless Steel Filter with Membrane 214848SP
Vaisala HUMICAP
®
180R (general purpose) HUMICAP180R
PT100 sensor 10429SP
Technical Support
For technical questions, contact the Vaisala technical support:
E-mail helpdesk@vaisala.com
Fax +358 9 8949 2790