Chapter 4 ________________________________________________________________ Operation
VAISALA_______________________________________________________________________ 37
0507-046
Figure 15 HyperTerminal Settings in Windows
®
2000
Environment
List of Commands
The text in BOLD letters in the [brackets] indicates the default setting.
Issue the commands by typing them on your computer keyboard. <cr>
stands for pressing Enter (on your computer keyboard). This section
presents the command list, the commands are described in detail later in
this chapter.
Measurement output
R Start continuous outputting
S Stop continuous outputting
INTV [0 … 255 S/MIN/H] Set continuous output interval for
RUN-mode
SEND [0 … 99] Output reading once
SMODE [STOP/RUN/POLL] Set serial interface mode
SERI [baud p d s] Serial line settings (default: 4800 E 7
1) baud: 300…19200
ADDR [0 … 99] Set transmitter address for Use in
POLL-mode
OPEN [0 … 99] Temporarily open transmitter from
POLL-mode to receive commands
CLOSE Set transmitter in POLL-mode

Table of Contents