Table 9 Output Quantities and their Metric and Nonmetric
Units
Quantity Metric Unit Nonmetric Unit
RH Relative Humidity %RH %RH
T Temperature
C F
TDF Dewpoint/frostpoint
C F
TD Dewpoint
C F
A Absolute humidity g/m
3
gr/ft
3
X Mixing ratio g/kg gr/lb
TW Wet bulb temperature
C F
PPM Humid air volume/dry air
volume
Ppm ppm
PW Water vapor pressure hPa lb/in
2
PWS Saturation water vapor
pressure
hPa lb/in
2
H Enthalpy kJ/kg Btu/lb
NOTE
Only the quantities selected when ordering the device can be selected for
an output quantity.
Examples:
>asel td tdf<cr>
Ch1 Td lo : -40.00 'C ?
Ch1 Td hi : 100.00 'C ?
Ch2 Tdf lo : -40.00 'C ?
Ch2 Tdf hi : 60.00 'C ?
>asel x td<cr>
Ch1 x lo : 0.00 g/kg ?
Ch1 x hi : 160.00 g/kg ?
Ch2 Td lo : -40.00 'C ?
Ch2 Td hi : 60.00 'C ?
>
ASCL Scale Analog Outputs
Syntax: ASCL<cr>
Example:
>ascl<cr>
Ch1 Td lo : -40.00 'C ?
Ch1 Td hi : 100.00 'C ?
Ch2 x lo : 0.00 g/kg ?
Ch2 x hi : 500.00 g/kg ?