To make a 2-point adjustment (GMP252 ppmCO
2
e
xample):
1. Connect to the wireless
configuration interface and open the Calibration menu.
2. Start the calibration mode with the Start calibration button.
If you cannot enter
configurations after selecting Start calibration, check that the
calibration PIN code is in place in the Settings > Probe menu.
3. If you need to set environmental compensations, enable and set the required
c
ompensations on the
Configuration tab.
Note that the environmental compensations you set on the Configuration tab and
the c
ompensations you set in the Settings > Probe menu are interconnected: the
configuration set in either menu is applied to both.
4. Remove any possible previous adjustments by restoring the factory adjustment: select
R
estore factory adjustment for each parameter you are adjusting.
5. Enter the calibration date and calibration information into the corresponding text fields.
6. Place the probe in the
first reference environment (adjustment point 1) and wait until
the measurement has stabilized. You can follow the stabilization from the
Measurements tab.
7. Enter the value of the first reference (for example, 0 if calibr
ating with a 0 ppmCO
2
reference gas) into the Reference value, point 1 field.
Chapter 7 – Calibration and Adjustment
53