EasyManua.ls Logo

Valmet OC2R - Creating and Performing Calibration

Default Icon
49 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Two-point calibration
Two-point calibration uses the same calculation formula as single-point calibration, and
coefficients A and B are calculated from the selected laboratory points. Two calibration
points are needed.
IMPORTANT: In two-point and multi-point calibration the difference between the
consistency points must be at least 25 % when calculated from the lowest point.
This helps to ensure that any errors in sampling or consistency analysis will not have a
noticeable effect on the slope of the curve. The device will alert the operator if the points
selected for calibration differ from each other by less than 25 %.
Multi-point calibration
The data set used for multi-point calibration should cover a wide range of ash and
consistency variations. Only samples for which a laboratory consistency result exists
can be used for calibration. Multi-point calibration requires at least 5 samples, and one
of the samples must differ from the rest by at least 25%.
10.4 Creating a recipe
In the Main result display press Esc.
Select 2 Calibration.
Select 1 Recipe.
Select a recipe, 1–4. In this example recipe 1 is
in use.
Press Edit/Save to start editing.
Calibration | 35 (48)
K20798 V1.1 EN - 1 - 12/2019

Related product manuals