ELEC T OUT OF LIMITS: Electronics Temperature is over the measurement limit.
Check the maximum value: Main menu 3=Diagnostics > 2=Diag. limits.
CLOCK NOT SET: TCU clock time not checked, still in the factory time zone. Set
the clock: Main menu 1=Configuration > 6=Set clock.
ANALOG OUTPUT1 SATURATED: Current output1 saturated. Check the "LRV"
and "URL" values (Main menu 1=Configuration > 2=Current output1).
ANALOG OUTPUT1 FIXED: Fixed current value set to current output.
ANALOG OUTPUT2 SATURATED: Current output2 saturated. Check the "LRV"
and "URL" values: Main menu 1=Configuration > 3=Current output2.
ANALOG OUTPUT2 FIXED: Fixed current value set to current output.
LED CURRENT WARNING: LED current is at the warning limit. Check the value:
Main menu 3=Diagnostics > 2=Diag.limits. Preventive maintenance is
recommended.
MEAS1 OUT OF LIMITS: The measured value is outside the measuring limits.
Check the limits: Main menu 3=Diagnostics > 2=Diag.limits.
MEAS2 OUT OF LIMITS: The measured value is outside the measuring limits.
Check the limits: Main menu 3=Diagnostics > 2=Diag.limits.
VCOMP OUT OF LIMITS: Reference signal is outside the measuring limits. Check
the limits: Main menu 3=Diagnostics > 2=Diag.limits.
Diagnostics | 45 (48)