Disabling the Packing Mode
Unless the packing mode disabled, no operation will happen even
after the equipment is turned on.
PaX-i Plus_Insight has a unique feature─ packing mode─ built in
the system to prevent the unit from being damaged while shipping
and transporting. Thus, it is in the packing mode by factory default.
The unit is required to exit the packing mode at this step for
successful installation.
1. Click General tab in the control panel.
2. In the Networking option, select Serial checkbox and set Serial port and baud
rateas follows and then click Manager.
If an error has occurred during connection, make sure the COM port
setting is correct as follows:
1) Run Device Manager.
2) Check the COM port assigned to virtual serial port under
com0com – serial port emulators.
3) If necessary, change the COM port in the Networking option and
click Save.
3. Enter the command PVER] to verify the current mode. Note that the equipment is now in
packing mode.