2. Choosing an Installation Site
16 PaX-i Plus / Insight Installation Manual
To assure line voltage quality, a separate 3-core grounded
power cable connected directly to central distribution panel with
over-current circuit breaker rated for 20A must be used.
Maximally allowed deviation of the tube voltage / tube current /
exposure time:
Tube Voltage (kVp) ± 10 % / Tube Current (mA) ± 20 % /
Exposure Time (s) ± (5 % + 50 ms) according to IEC 60601-2-
63.
The mains resistance should not exceed 0.045 ohm at 100 V
and 0.19 ohm at 240 V. This equipment should be connected to
the earthed outlet.