8. Acquiring CT Images
102
I
PaX-i3D Smart
Getting prepared
1. Ask the patient to remove all the metal objects (glasses, earrings, hair pins, braces,
false teeth). Metal objects may induce ghost images and lower the image quality.
2. Ask the patient to wear a lead apron to protect them from residual radiation.
3. Use the Vertical Frame Up/Down button or switch option to adjust the
equipment to match the height of the patient.
Patient Positioning
1. Insert the bite block into the chin rest then cover with a sanitary vinyl cover.
Bite Block
Bite Block Lock/Unlock Knob
Sanitary Vinyl Cover
The sanitary vinyl cover is for single use only. It should be
replaced after each patient. Be sure to use the approved vinyl
cover.
Clean the chin rest and bite block with ethanol and wipe with a dry towel
before the next patient.