PaX-i3D Green (PHT-60CFO) User Manual
158
Appendices
Guidance and manufacturer’s declaration - electromagnetic immunity
The PaX-i3D Green (Model: PHT-60CFO) is intended for use in the
electromagnetic environment specied below.
The customer or the user of the PaX-i3D Green (Model: PHT-60CFO)
should assure that it is used in such an environment.
Immunity
Test
IEC 60601-1-2
Test level
Compliance
level
Electromagnetic environment
-guidance
Conducted
RF
IEC
61000-4-6
3 Vrms
150 kHz to
80MHz V1=3Vrms
Portable and mobile RF
communications equipment
should be used no closer to
any part of the PaX-i3D Green
(Model: PHT-60CFO), including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance:
= [ ]
3,5
V
1
= [ ]
3,5
E
1
80 MHz to 800MHz
= [ ]
7
E
1
800 MHz to 2,5GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from xed RF
transmitters, as determined by
an electromagnetic site survey,
a
should be less than the
compliance level in each
frequency range.
b
Interference may occur in the
vicinity of the equipment marked
with the following symbol :
Radiated
RF
IEC
61000-4-3
3 V/m
80 MHz to
2.5GHz
E1=3V/m