43
Operations & Maintenance Manual: Cleaning & Disinfecting
PRODUCT
DISINFECTION
LEVEL
CONDITIONS
Isopropyl alcohol solution
(70%)
Low As per chemical manufacturer’s instructions
Oxivir
®
Tb Wipes Low Up to 1100 cycles as per chemical manufacturer’s instructions
Sani‑Cloth
®
Bleach Wipes Low Up to 750 cycles as per chemical manufacturer’s instructions
SaniCloth
®
AF3 Germicidal
Wipes
Cleaner Up to 2000 cycles as per chemical manufacturer’s instructions
Low
Exposure: Using fresh wipe(s), wet all surfaces and allow to remain
wet for 3 minutes.
Rinse: Not applicable. Allow the component to thoroughly air dry.
ASP
®
CIDEXPLUS
®
28Day
Solution
High As per chemical manufacturer’s instructions
ASP
®
CIDEX
®
OPA High As per chemical manufacturer’s instructions
Metrex
®
MetriCide
®
Plus30
High As per chemical manufacturer’s instructions
Medivators
®
Rapicide
®
High
100 cycles in a Medivators Advantage
®
Plus automated endoscope
reprocessor (AER) disinfection routine meeting the following
requirements:
• Concentration of disinfectant – 750‑950 parts per million
• Temperature – 28°C–32°C (82.4°F – 89.6°F)
• Exposure time – 5 minutes
• AER configuration – Hookup 2‑8‑002HAN rev A
• AER parameter set – 1‑35‑101 C DISF
Sultan
®
Healthcare
Sporox
®
II
High As per chemical manufacturer’s instructions
STERIS
®
S40™
or S20™
Sterilization
Standard cycles in the following processors:
STERIS
®
SYSTEM 1
®
(outside U.S.)
SYSTEM 1E
®
(in U.S.)
SYSTEM 1 EXPRESS (outside U.S.)
SYSTEM 1 PLUS (outside U.S.)
STERIS
®
V‑PRO
®
low
temperature sterilization
systems
Sterilization Up to 200 cycles as per manufacturer’s instructions
ASP
®
Hydrogen
PeroxideGas Plasma
Sterilization
STERRAD
®
100S (in U.S.),
STERRAD
®
100S short cycle (outside U.S.),
STERRAD
®
NX standard cycle, or
STERRAD
®
100NX standard cycle