5.6. WEIGHT
H2 FLOW RATE CONFIGURATION AIR FLOW RATE NM PLUS FID STATION
100 mL/min 1800/5000 41 kg (90.4 lbs)
300 mL/min 1800/5000 42 kg (92.6 lbs)
600 mL/min 1800/5000 43 kg (94.8 lbs)
1000 mL/min 1800/5000 44 kg (97 lbs)
H2 FLOW RATE CONFIGURATION AIR FLOW RATE PG PLUS FID STATION
100 mL/min 1800/5000 39 kg (86 lbs)
250 mL/min 1800/5000 39 kg (86 lbs)
5.7. TRANSPORT AND STORAGE
Temperature Minimum -15ºC (59ºF), Maximum 35ºC (95ºF)
Humidity Less than 90% RH, non-condensing
Shocks and Vibrations during transport Shocks <1g; Vibrations <0.2g/5Hz
Instrument position Horizontal
The above mentioned values are considered for packing.
Recommendations for Transport:
If the instrument needs to be shipped back to the factory for repair, the gas generator should be packed in its
original packaging with the inlets and outlets protected against any accidental
exposure to dust or particles.
Purge water before moving the instrument. Do not move when it is full of water.