22
NM & PG PLUS FID STATIONS
USER MANUAL
Version 1.03.0000
www.vicidbs.com
At this point, make sure that you have connected the hydrogen outlet
to your GC or application. Once done, press Open.
The hydrogen will go through the outlet and the pressure will rise
until it reaches the selected pressure.
Once reached, the generator will show “Normal Flow”
6.7. TURNING OFF THE GENERATOR
To stop the Plus FID Station hydrogen component, press the
H
2
ON/OFF button. Button light will turn off.
6.8. STARTING THE ZERO AIR GENERATOR
The zero air generator needs to be connected to an external compressed air source. In order to protect the
generator, it is necessary to place a specic dust lter at the air inlet of the generator (delivered with the
generator).