NM & PG PLUS FID STATIONS
USER MANUAL
Version 1.03.0000
www.vicidbs.com
8.3. CONFIGURE PARAMETERS
SELECTION DESCRIPTION OPTIONS DEFAULT
Set Clock Adjust the internal clock - -
Low Press Error* Percentage of total pressure that has to be
exceeded to activate the low pressure alarm
0-100% 30%
Pressure Rise Sets how fast the pressure should increase.
If the pressure increases at a slower rate, the
low pressure alarm will activate.
0.01 to 4.2 bar/min 14
0.2 to 60 psi/min
0.09 bar
14 psi
Pressure Drop
Delay
Sets a delay in seconds to ignore a pressure
drop (override low pressure alarm)
2 to 10 minutes 2 minutes
Auto Start After powering the generator, it goes back to
the status before power shutdown.
Exam ple:
1. System running, producing hydrogen
and valve is open
2. Power shutdown
3. Power restarted
4. System starts running and then opens
the valves automatically after reaching
the internal pressure
YES / NO NO
Beeper Sets whether the audible signal is activated
in the event of an alarm
ON / OFF ON
Cascading Mode Sets the cascading mode. (see cascading for
more details)
None / Normal /
Ext. P control
None
N Gen H2 in
Cascading
Sets the number of generators in the
cascading chain
1 to 10 1
ID Number Sets the ID number 1 to 10 1
Remote Start/Stop
Mode
Congures the remote Start/Stop function Start/Stop, Start only,
Direct control
Start/Stop
Auto Rell If set to ON, the pre-level water alarm is
used to trigger an external pump or valve to
rell the water tank
ON / OFF OFF
Auto Rell Timeout Within this time the internal water tank
must be lled. If a time out occurs the alarm
“check auto rell” appears.
5-600 sec. 30 sec.
Remote Relay
Mode
Congures the remote relay contacts Normally Open (NO)
Normally Closed (NC)
NC
Pre Alarms in List If set to YES, the pre alarms are also shown
in the alarm log
YES / NO NO
Pressure Units Sets the desired unit of measure for the
pressure
bar / psi / kPa bar