Installation Site Requirements 1
1
Chapter 1
Installation
The Waters
®
2996 Photodiode Array (PDA) Detector operates in any standard laboratory
environment. The detector requires electrical power, sample and waste fluidic lines, and
either the Millennium®
32
or MassLynx™ workstations. Optional connections on the
detector rear panel allow communication with chart recorders, data integrators, and other
instruments that are not compatible with Millennium
32
or MassLynx software control.
1.1 Installation Site Requirements
Install the 2996 PDA Detector at a site that meets the specifications in Table 1-1 and
Figure 1-1.
Table 1-1 Site Requirements
Factor Specification
Ambient temperature 4 to 40 °C (39 to 104 °F), not to vary more
than 1 °C per hour (to prevent drift)
Relative humidity 20 to 80%, noncondensing
Bench space Width: 11.5 in. (29 cm)
Depth: 24 in. (61 cm)
Height: 8.5 in. (22 cm)
Bench support Capable of supporting 31.5 pounds (14.3 kg)
Clearance At least 4 in. (10 cm) on the back side for
ventilation
Power Grounded ac, 100 to 240 Vac