EasyManua.ls Logo

Waters 2996 - Millennium³² Workstation Connections

Waters 2996
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Millennium
32
Workstation Connections 3
1
the Waters 2996 Detector Startup Kit) in the fuse holder in the rear of the detector (refer to
Section 3.3, Replacing the Fuses).
Connecting the Power Cord
Connect one end of the 2996 Detector power cord to the rear panel power receptacle
(Figure 1-2) and the other end to a power outlet.
Figure 1-2 Detector Rear Panel
1.3 Millennium
32
Workstation Connections
The 2996 Detector requires signal connections to the Millennium
32
Workstation over the
IEEE-488 bus. All detector control and data acquisition communications take place over
the IEEE-488 bus.
Note: If an inject start signal is not available over the IEEE-488 bus, you must provide a
signal at the Event In 1 terminals on the 2996 Detector rear panel (Section 1.4.2,
Connecting Event Cables).
TP01452
AC Input

Table of Contents

Related product manuals