EasyManua.ls Logo

Waters 2996 - Chapter 3 Maintenance; Flow Cell Maintenance

Waters 2996
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Flow Cell Maintenance 19
3
Chapter 3
Maintenance
This chapter covers maintenance of the Waters 2996 Photodiode Array Detector flow cell,
lamp, and fuse.
3.1 Flow Cell Maintenance
The flow cell requires maintenance when:
The reference spectrum changes.
The cell fluid leaks out of the drain tube.
The Lamp diagnostic (in the Millennium
32
PDA Diagnostics window) fails and the
lamp status light is on (Table 2-1).
The 2996 Detector causes high backpressure.
Note: Conditions other than a dirty flow cell may cause decreased lamp intensity. For
more information, refer to Chapter 2, Diagnostics and Calibration.
Flow cell maintenance consists of:
Flushing the flow cell
Removing the flow cell
Disassembling and cleaning the flow cell
Installing the flow cell assembly
Caution: To avoid possible electric shock, do not remove the 2996 Detector power
supply covers. The power supply does not contain user-serviceable components.

Table of Contents

Related product manuals