EasyManua.ls Logo

Waters 2996 - Chapter 4 Principles of the 2996 PDA Detector Optics; 2996 Detector Optics

Waters 2996
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2996 Detector Optics 29
4
Chapter 4
Principles of the 2996 PDA
Detector Optics
To use the Millennium
32
PDA software effectively, you must be familiar with the
principles of operation of the optics and electronics of the Waters 2996 PDA Detector.
4.1 2996 Detector Optics
The 2996 Detector is an ultraviolet/visible light (UV/Vis) spectrophotometer with:
512 photodiodes
Optical resolution of 1.2 nm per diode
Operating wavelength range from 190 to 800 nm
The light path through the optics assembly of the 2996 Detector is shown in Figure 4-1.