2-4 Installing the Waters 515 Pump
5. Verify that the serial number on the pump matches the serial number
on the Certificate of Validation. Keep the Certificate of Validation with
this guide for future reference.
6. Inspect all items for damage. Immediately report any shipping damage
to both the shipping company and your Waters representative. Contact
Waters Technical Service at 800 252-4752, Canadian and U.S.
customers only. Other customers, call your local Waters subsidiary or
call Waters corporate headquarters for assistance in Milford,
Massachusetts (U.S.A.). See the Waters Licenses, Warranties, and
Support document for complete information on reporting shipping
damages and submitting claims.
Making electrical connections
For proper operation, the 515 HPLC Pump requires:
•Installed fuses
A grounded AC power supply with no abrupt voltage fluctuations
A connection to an external controller or pump control module (for
remote operation only)
Required materials
Small flat-blade screwdriver
Fuses (if replacing fuses)
Pump control cable from external controller (if connecting for remote
operation)
Power cord (Startup Kit)
Replacing the power supply fuses
The 515 pump is shipped with two 3.15 A fuses installed.
Warning: To avoid electrical shock and possible injury, remove the
power cord from the rear panel of the instrument before you perform the
procedure in this section.