1-4 Waters 515 Pump Overview
Electronics components
Before you install the 515 pump, familiarize yourself with its electronics
components as illustrated in the figure below. The table titled “Electronics
components” on page 1-5 describes the functions of the electronics
components.
Inlet and outlet check
valve assemblies
Maintain flow direction by opening in one
direction only.
High-pressure noise
filter
Dampens operating pressure fluctuations.
Inlet manifold Provides connection for eluent inlet tubing and
routes eluent to the inlet check valve on each
pump head.
Pressure transducer Senses operating pressure and converts value to
electronic signal for monitoring.
Pump outlet Routes eluent to injector, column, and detector.
Fluid-handling components (Continued)
Component Function
Caution: To prevent eluent from leaking
out of the pump outlet, make sure you
position the eluent reservoir below the
pump inlet manifold before you
disconnect the pump outlet fitting.