1 ACQUITY UPLC I-Class IVD System
18 July 7, 2014, 715003736IVD Rev. C
•ACQUITY UPLC I-Class Binary Solvent Manager (BSM)
•ACQUITY UPLC I-Class Sample Manager - Flow-Through-Needle (SM -
FTN)
•50 µL extension loop for SM-FTN
•Column Heater with active pre-heating (CH-A) and tubing
•Column Heater (CH-A) extension kit for mass spectrometer
•Xevo TQ-S IVD mass spectrometer
• ACQUITY UPLC I-Class System Startup kit, including solvent tray
module
•Leak sensors
•ACQUITY UPLC BEH C
18
column, 1.7 µm, 2.1 x 50 mm, IVD
•ACQUITY UPLC I-Class IVD system documentation CD, part number
715003740IVD
• Workstation and MassLynx, ACQUITY UPLC Console software, and
Connections INSIGHT software
•FlexCart, optional
Binary solvent manager
The Waters ACQUITY UPLC I-Class IVD Binary Solvent Manager (BSM)
delivers solvent compositions for isocratic and binary gradient methods at
flow rates of between 0.01 and 2.0 mL per minute. Its features include in-line
filters upstream of a primary check valve, the Waters
®
Intelligent Intake
Valve (i
2
Valve), automated priming functions, and daily system-setup
routines.
The design of the binary solvent manager is optimized for sub-2-µm particle
liquid chromatography. The upper pressure limit is set at 124,106 kPa
(1241 bar, 18,000 psi) per pump. A maximum flow rate of 1 mL/min is
permitted at system backpressures to 124,106 kPa (1241 bar, 18,000 psi),
which falls within the range of optimal linear velocity for sub-2-µm particle
columns of ID 1.0 mm to 3.0 mm.
See also: The ACQUITY UPLC I-Class IVD Binary Solvent Manager
Overview and Maintenance Guide (part number 715003739IVD) for more
details.