Preparing system hardware
July 7, 2014, 715003736IVD Rev. C 31
Run LED (sample manager)
The run LED, on the right-hand side of the power LED on the sample
manager’s front panel, indicates the run status. A steady-green run LED
indicates that injections are being run.
Enabling the leak sensors
Rule: When you power-on the system, the leak sensors default to disabled
status unless previously enabled.
To enable the leak sensors:
1. In the ACQUITY UPLC Console, select Control > Leak Sensors.
Status LED indications:
LED mode and color Description
Unlit Binary solvent manager and sample manager –
Indicates the device is currently idle.
Steady green Binary solvent manager – Indicates solvent is
flowing.
•Sample manager Indicates the sample
manager is operating normally, attempting to
complete any outstanding samples or
diagnostic function requests. When sample
and diagnostic function requests are finished,
the LED reverts to the unlit mode.
Flashing green Binary solvent manager and sample manager –
Indicates that the module is initializing.
Flashing red Indicates that an error stopped the instrument or
device. Refer to the ACQUITY UPLC Console for
information regarding the error.
Steady red Indicates a module’s failure that prevents its
further operation. Power-off the module, and
then power-on. If the LED is still steady red,
contact your Waters service representative.