41
WIKA operating instructions, model CEP6000
MM/YYYY country code based on 10/2014 GB/CN
GB
CAUTION!
Damage to property
Improper cleaning may lead to damage to the instrument!
Do not use aggressive cleaning agents.
Do not use any pointed and hard objects for cleaning.
Electrical connections must not come into contact with moisture.
8.3 Recalibration
DKD/DAkkS certificate - official certificates:
We recommend that the instrument is regularly recalibrated by the manufacturer, with
time intervals of approx. 12 months. The basic settings will be corrected if necessary.
9. Dismounting, return and disposal
Personnel: Skilled personnel
9.1 Dismounting
WARNING!
Physical injury
When dismounting the optional external pressure sensor, there is a
danger from aggressive media and high pressures.
Observe the information in the material safety data sheet for the
corresponding medium.
Disconnect the pressure sensor once the system has been
depressurised.
9.2 Return
Strictly observe the following when shipping the instrument:
All instruments delivered to WIKA must be free from any kind of hazardous substances
(acids, bases, solutions, etc.) and must therefore be cleaned before being returned.
WARNING!
Physical injuries and damage to property and the environment
through residual media
Residual media on the model CEP6000 hand-held multi-function calibrator
can result in a risk to persons, the environment and equipment.
With hazardous substances, include the material safety data sheet for
the corresponding medium.
Clean the instrument, see chapter 8.2 “Cleaning”.