25
WIKA operating instructions, model CEP6000
MM/YYYY country code based on 10/2014 GB/CN
GB
MEASURE / SOURCE
3W
mA+
TC
V
mA
Loop
+
V
Hz
+
4W
mA–
MEASURE
+
+
+
+
+
+
3W
4W
RTD
RT
D
RT
D
2W
mA IN
0.000 mA
-25.00%
RTD IN 4W P10-385
MENU
25.0 °C
LIGHT
Measuring the temperature with RTD probe connected
4. Select the corresponding RTD type in the menu.
5. Select temperature unit.
6. Connect the RTD wires (see following figure).
The resistance can also be measured with this function. For this, follow the steps as
described above, and as RTD type, select “OHMS”. With this option and a measure-
ment table, an RTD probe that is not programmed into the calibrator can also be used for
measurement.
6.4.4 Measuring pressure
CAUTION!
Pressurisation
Damage within the pressure system due to open valves.
1. Close the valve.
2. Release the pressure slowly.
3. Connect the pressure module to the pressure line.
ATTENTION!
Damage to the threaded connections
The pressure module can be damaged by over-tightening the threaded
connection.
Use suitable tools.
Tighten the pressure connection with a maximum of 13.6 Nm (torque
spanner).