pH/Cond 3320 Ion concentration
ba77054e03 03/2018 41
7. Thoroughly rinse the electrode with distilled water.
8. Immerse the electrode in standard solution 1.
9. When calibrating without temperature sensor:
Measure the temperature of the standard solution using a thermom-
eter.
Use <F2>/[ ⇅ ] to select the setting of the temperature.
Use < >< > to set the temperature.
Use <F2>/[ ⇅ ] to select the setting of the concentration.
10. Set the concentration of the standard solution with < >< > and press
<ENTER>.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
11. Wait for the end of the AutoRead measurement or accept the calibration
value with <ENTER>.
The calibration display for the next standard solution appears.