Ion concentration pH/Cond 3320
42 ba77054e03 03/2018
Continuing with two-
point calibration
ISE
0.0
24.8 °C
Ag 0.020 mg/l
2
15.02.2014
12:12
Standard
mV
12. Thoroughly rinse the electrode with distilled water.
13. Immerse the electrode in standard solution 2.
14. When calibrating without temperature sensor:
Measure the temperature of the standard solution using a thermom-
eter.
Use <F2>/[ ⇅ ] to select the setting of the temperature.
Use < >< > to set the temperature.
Use <F2>/[ ⇅ ] to select the setting of the concentration.
15. Set the concentration of the standard solution with < >< > and
press <ENTER>.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
16. Wait for the end of the AutoRead measurement or accept the calibra-
tion value with <ENTER>.
The calibration display for the next standard solution appears.
ISE
8.9
24.8 °C
15.02.2014
12:12
0.02 mg/l
mV
AR
ISE
20.5
24.8 °C
Ag 0.050 mg/l
3
15.02.2014
12:12
Standard
mV