Ci7XX0 BENCHTOP SPECTROPHOTOMETER
12
Lens and Aperture Plate Indicators (*, 3.5, 6, 10, 17, 25)
The top row of LEDs illuminates to indicate the current lens position. The “*” illuminates
when the lens is in a custom position. The LED will light red if the unit is configured with a
lens position larger than the installed aperture plate size.
The bottom row of LEDs illuminates to indicate the currently installed aperture plate. The
“*” illuminates for custom plates and transmission aperture plates (with white reflective
ring), and blinks if the instrument could not detect the installed plate. If auto is enabled
for the lens, then the installation of an aperture plate automatically sets the lens to match
the aperture plate.
Standard Button and Indicator
The standard button is used to initiate a standard measurement. Note: The button must
be supported by your application software and function as programmed.
The indicator illuminates green when programmed to perform a measurement. It may
also be programmed to flash green to support features in your application.
Preview Button and Indicator
The preview button is used in conjunction with the external video monitor preview
feature.
The indicator illuminates green when it is activated.
The video preview is toggled when a reflectance measurement mode is selected. When a
transmission measurement mode is selected, the button will toggle the green
transmission targeting laser on and off.
The behavior of the button may also be programmed in your software application.
Trial Button and Indicator
The trial button is used to initiate a trial measurement. Note: The button must be
supported by your software application and function as programmed.
The indicator illuminates green when programmed to perform a measurement. It may
also be programmed to flash green to support features in your application.
NetProfiler
Indicator Off: NetProfiler feature is not enabled.
Solid Green: NetProfiler subscription is currently activated.
Solid Amber: The profile has expired and updating is recommended.
Calibration
Solid Red: Calibration is required
Solid Green: White, black, and UV calibration are not required at this time
Solid Amber: White and black calibration are not needed, but UV calibration has expired.
Measurements can still be taken in this condition. It is up to the user to decide if the UV
calibration should be updated at this time.
Transform
Indicator Off: Transform feature in not enabled
Solid Green: Transform feature is activated
Specular Included (SCI)
Illuminates when a specular included measurement is selected.