Remote Operation
GPIB Operation
122 Operating Manual for XDC Series Power Supply
SCPI
To set up GPIB control parameters:
SYST:COMM:GPIB:ADDR <GPIB-address>
where
• the GPIB address may be in the range 1 to 30.
To configure the unit to generate a power-on service request:
SYST:COMM:GPIB:PONS {ON | OFF}
To select GPIB as the remote control source:
SYST:REM:SOUR GPIB
Using GPIB Sending a GPIB command should put the power supply in remote mode with the
RMT annunciator lit.
Refer to the manual for your GPIB interface card. Commands to change remote and
local mode will be specific to that card.
Press the LCL/RMT button to return to local mode, except if the power supply is in
LLO (local lockout) state.