EasyManuals Logo

Xantrex XDC 40-150 User Manual

Xantrex XDC 40-150
256 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #135 background imageLoading...
Page #135 background image
Remote Operation
SCPI Commands for Digital Interfaces
132 Operating Manual for XDC Series Power Supply
Configure
Other
Protection
Mechanisms
Fold Protection
Fold protection causes the supply to shut down if the selected regulation mode is
entered. A delay time may be specified as well.
To set the fold mode:
OUTP:PROT:FOLD {CC|CV|CP|NONE}
Where:
NONE indicates fold protection is disabled.
CC indicates the supply will shut down due to constant current condition.
CV indicates the supply will shut down due to constant voltage.
CP indicates the supply will shut down due to constant power.
The default value is NONE.
To set the fold delay:
OUTP:PROT:FOLD:DEL <delay_time>
Where:
<delay_time> is a value in the range of 0–60 seconds.
Units may be specified as “ms” (milliseconds) “s” (seconds), or “min” (minutes). If
no units are specified, the default seconds are assumed. (Increments of 0.1s are
allowed.) The default value is 0.5 second.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Xantrex XDC 40-150 and is the answer not in the manual?

Xantrex XDC 40-150 Specifications

General IconGeneral
BrandXantrex
ModelXDC 40-150
CategoryPower Supply
LanguageEnglish

Related product manuals