2.4 Voltage Source Corresponding to TC Thermoelectromotive Force
2-8
IM CA500-01EN
9.
Using the arrow keys, select TC-A to use the TC-A terminal (thermocouple mini plug) or TC-B
to use TC-B.
To finish entering the settings here, proceed to step 14.
Setting the Reference Junction Compensation (RJC) (when using the TC-B
terminal)
10.
Select TC-B RJC. ON and OFF appear in the selection menu.
11.
Use the arrow keys to set RJC to ON or OFF.
To finish entering the settings here, proceed to step 14.
Setting the Temperature Scale
12.
Select Temperature Scale. Options appear in the selection menu.
13.
Use the arrow keys to set the temperature scale.
Confirming the Settings
14.
Press the arrow key corresponding to EXIT SETUP. The settings are confirmed, and a screen
appears showing the source value and measurement value.
The cancel the settings, press ESC to return to the screen in step 6.
To initialize the settings, pressing the arrow key corresponding to INIT SETUP.
Turning the Source On and Off
15.
With the source value and measurement value displayed, press OUTPUT ON/OFF.
The source value set in step five is output, and OUTPUT:OFF on the screen changes to
OUTPUT:ON.
To turn off the source, press OUTPUT ON/OFF again.
Divided Source
See section 2.8, “Dividing and Generating the Source Values”.
Sourcing with the Sweep Function
See section 2.9, “Sweep Source”.