Confirming the Contact Output
17.
Press the arrow key corresponding to EXIT SETUP. The settings are confirmed, and the screen
reverts to show the source value and measurement value.
The cancel the settings, press ESC to return to the screen and step 14.
To initialize the settings, pressing the arrow key corresponding to INIT SETUP.
Turning the Source On and Off
18.
With the source value and measurement value displayed, press OUTPUT ON/OFF.
The source value set in step 5 is output. The signal is generated according to the specified
pulse count.
To turn off the source, press OUTPUT ON/OFF again.
Divided Source
See section 2.8, “Dividing and Generating the Source Values”.
2.6 Frequency and Pulse Source
2-14
IM CA500-01EN