Page #102 background imageLoading...
Page #102 background image
5-68
IM 701310-17E
:CHANnel<x>:PROBe:AUTO?
Function Queries the probe attenuation of the channel when
set to AUTO.
Syntax
:CHANnel<x>:PROBe:AUTO?
<x> = 1 to 4
Example
:CHANNEL1:PROBE:AUTO? -> :CHANNEL1:
PROBE:AUTO 1
:CHANnel<x>:PZCalibrate[:EXECute]
Function Execute the zero adjustment of the current probe of
the channel.
Syntax :CHANnel<x>:PZCalibrate[:EXECute]
<x> = 1 to 4
Example :CHANNEL1:PZCALIBRATE:EXECUTE
:CHANnel<x>:SELect
Function Sets the waveform (input/computation) to be
assigned to the input channel or queries the current
setting.
Syntax
:CHANnel<x>:SELect {INPut|MATH}
:CHANnel<x>:SELect?
<x> = 1 to 4
Example
:CHANNEL1:SELECT INPUT
:CHANNEL1:SELECT? -> :CHANNEL1:SELECT
INPUT
:CHANnel<x>:SVALue (Scale VALUE)
Function Turns ON/OFF the scale display of the channel or
queries the current setting.
Syntax
:CHANnel<x>:SVALue {<Boolean>}
:CHANnel<x>:SVALue?
<x> = 1 to 4
Example
:CHANNEL1:SVALUE ON
:CHANNEL1:SVALUE? -> :CHANNEL1:SVALUE 1
:CHANnel<x>:VDIV
Function Sets the vertical sensitivity (V/div) of the channel or
queries the current setting.
Syntax
:CHANnel<x>:VDIV {<Voltage>|<Current>}
:CHANnel<x>:VDIV?
<x> = 1 to 4
<V
oltage> and <Current> = See the DL9000 User’s
Manual.
Example
:CHANNEL1:VDIV 5V
:CHANNEL1:VDIV? -> :CHANNEL1:
VDIV 5.000E+00
:CHANnel<x>:OCANcel
Function Turns ON/OFF the offset cancel of the channel or
queries the current setting.
Syntax
:CHANneL<x>:OCANcel {<Boolean>}
:CHANneL<x>:OCANcel?
<x> = 1 to 4
Example
:CHANNEL1:OCANCEL ON
:CHANNELL1:OCANCEL? -> :CHANNEL1:
OCANCEL 1
:CHANnel<x>:OFFSet
Function Sets the offset voltage of the channel or queries the
current setting.
Syntax
:CHANneL<x>:OFFSet
{<Voltage>|<Current>}
:CHANneL<x>:OFFSet?
<x> = 1 to 4
<V
oltage> and <Current> = See the DL9000 User’s
Manual.
Example
:CHANNEL1:OFFSET 1V
:CHANNEL1:OFFSET? -> :CHANNEL1:
OFFSET 1.000E+00
:CHANnel<x>:POSition
Function Sets the vertical position of the channel or queries the
current setting.
Syntax
:CHANneL<x>:POSition {<NRf>}
:CHANneL<x>:POSition?
<x> = 1 to 4
<NRf> = –4 to 4 (div)
Example
:CHANNEL1:POSITION 1
:CHANNEL1:POSITION? -> :CHANNEL1:
POSITION 1.000E+00
:CHANnel<x>:PROBe?
Function Queries all settings related to the probe attenuation
of the channel.
Syntax
:CHANneL<x>:PROBe?
<x> = 1 to 4
Example
:CHANNEL1:PROBE? -> :CHANNEL1:PROBE:
MODE 1
:CHANnel<x>:PROBe[:MODE]
Function Sets the probe attenuation of the channel or queries
the current setting.
Syntax
:CHANneL<x>:PROBe[:MODE] {<NRf>|AUTO|
C1|C10|C100}
:CHANneL<x>:PROBe[:MODE]?
<x> = 1 to 4
<NRf> = 1, 2, 5, 10, 20, 50, 100, 200, 500,or 1000
Example
:CHANNEL1:PROBE:MODE 1
:CHANNEL1:PROBE:MODE? -> :CHANNEL1:
PROBE:MODE 1
5.6 CHANnel Group