5-148
IM 701310-17E
:TIMebase?
Function Queries all settings related to the time base.
Syntax
:TIMebase?
Example
:TIMEBASE? -> :TIMEBASE:TDIV 1.000E-06
:TIMebase:SRATe? (Sample RATE)
Function Queries the sample rate or queries the current
setting.
Syntax
:TIMebase:SRATe?
Example
:TIMEBASE:SRATE? -> :TIMEBASE:
SRATE 125.0E+06
:TIMebase:TDIV
Function Sets the T/div value or queries the current setting.
Syntax
:TIMebase:TDIV {<Time>}
:TIMebase:TDIV?
<T
ime> = 500 ps to 50 s
Example
:TIMEBASE:TDIV 1NS
:TIMEBASE:TDIV? -> :TIMEBASE:
TDIV 1.000E-06
5.28 TIMebase Group

Table of Contents