Index
Error in Communication Execution (200 to 299)
Code Messages Corrective Action Reference Page
221 Setting conflict. Check the relevant settings. Chapter 5
222 Data out of range. Check the range. Chapter 5
223 Too much data. Check the length of the data. Chapter 5
224 Illegal parameter value. Check the range. Chapter 5
241 Hardware missing. Check the installed options.
260 Expression error. Equations cannot be used.
Error in Communication Query (400 to 499)
Code Messages Corrective Action Reference Page
410 Query INTERRUPTED. Check transmission/reception order. 4-2
420 Query UNTERMINATED. Check transmission/reception order. 4-2
430 Query DEADLOCKED. Limit the length of the program message including <PMT> to
1024 bytes or less.
4-2
440 Query UNTERMINATED after indefinite
response.
Do not specify a query after the *IDN? or *OPT? command.
Error in System Operation (399)
Code Messages Corrective Action Reference Page
399 Fatal error in the communication driver. Maintenance service is required.
Warning (50)
Code Messages Corrective Action Reference Page
50 *OPC/? exists in message. Place the *OPC or *OPC? command at the end of the
program message.
Other Errors (350)
Code Messages Corrective Action Reference Page
350 Queue overflow. Read the error queue. 6-6
Note
Code 350 indicates overflow of error queue. This code is returned as a response to the
STATus:ERRor? query; it does not appear on the screen.